FDA Adverse Event Malfunction Summary report: N

I-STAT CHEM8+ CARTRIDGE

MDR report key: 22435527 · Received July 8, 2025

Report

Report Number
2245578-2025-00075
Event Type
Malfunction
Date Received
July 8, 2025
Date of Event
June 13, 2025
Report Date
August 11, 2025
Manufacturer
ABBOTT POINT OF CARE
Product Code
JPI
PMA / PMN Number
K183680
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

APOC INCIDENT # (B)(4) APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.

Additional Manufacturer Narrative · 0

APOC INCIDENT: # 981707 THE INVESTIGATION WAS COMPLETED ON 06-AUG-2025. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) CONFIRMED THE CARTRIDGE LOT MET FINISHED GOODS RELEASE CRITERIA. RETAINED TESTING MET THE ACCEPTANCE CRITERIA FOUND IN Q04.01.003 REV. AO, APPENDIX 1 - PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE. NO DEFICIENCY HAS BEEN DETERMINED FOR CHEM8+ CARTRIDGE LOT H25079.

Description of Event or Problem · 0

ON (B)(6) 2025, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CHEM8+ CARTRIDGE THAT YIELDED A DISCREPANT CREATININE RESULT ON AN 78-YEAR-OLD MALE PATIENT. THERE WAS NO ADDITIONAL PATIENT INFORMATION AT THE TIME OF THIS REPORT. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. METHOD: DATE: TESTED NA K: GLU HCT SAMPLE: I-STAT (B)(6) 2025, 10:23 110 >9.0, 102, 21 WB, I-STAT (B)(6) 2025 , 10:37 112 >9.0, 97 , <15 WB, ROCHE (B)(6) 2025, 12:31 141 4.7, 111, 34.7 NI, I-STAT (B)(6) 2025 , 15:35 138 4.5 , 110, <15 PLAMA. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON THE LIMITED INFORMATION AVAILABLE THAT SUGGESTED THAT PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY. THE INVESTIGATION IS UNDERWAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
341508 I-STAT CHEM8+ CARTRIDGE CHEM8+ CARTRIDGE JPI ABBOTT POINT OF CARE NA H25079

Patients

Seq Age Sex Outcome Treatment
1 78 YR Male