FDA Adverse Event Malfunction Summary report: N

LIBERTY CYCLER SET, DUEL PATIENT CONNECT

MDR report key: 2243541 · Received August 19, 2011

Report

Report Number
8030665-2011-00090
Event Type
Malfunction
Date Received
August 19, 2011
Date of Event
February 1, 2011
Report Date
August 19, 2011
Manufacturer
REYNOSA MANUFACTURING
Product Code
FKX
PMA / PMN Number
K043363
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE HISTORY RECORD REVIEW: THE DHR OF THIS PRODUCT AND LOT WAS REVIEWED AND THE FOLLOWING WAS FOUND: THIS PRODUCT DID NOT HAVE ANY NCR. NO DEVIATION WAS DOCUMENTED DURING THE MANUFACTURING PROCESS. ALL THE APPLICABLE TESTS DURING THE IN PROCESS AND FINAL INSPECTIONS WERE CONDUCTED IN ORDER TO ENSURE PRODUCT INTEGRITY AND DOCUMENTED WITH ACCEPTABLE RESULTS ACCORDING TO APPLICABLE PROCEDURES. NO REWORKS/RE-INSPECTIONS WERE PERFORMED TO THIS FINISH GOOD LOT 10HR08094. THE SAMPLE WAS NOT RECEIVED, THEREFORE, THE COMPLAINT IS DEEMED AS UNCONFIRMED. (B)(4) WILL CONTINUE MONITORING THIS TYPE OF INCIDENT PER CURRENT PROCEDURES.

Description of Event or Problem · 1

A PERITONEAL DIALYSIS NURSE CALLED STATING THE CASSETTE HAD LEAKED FLUID INSIDE OF CASSETTE DOOR. THERE IS NO REPORT OF PT ILL EFFECT. THE SAMPLE IS NOT AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIBERTY CYCLER SET, DUEL PATIENT CONNECT PERITONEAL DIALYSIS TUBING SET FKX REYNOSA MANUFACTURING NA 10HR08094

Patients

Seq Age Sex Outcome Treatment
1 NI LIBERTY DIALYSIS MACHINE