FDA Adverse Event
Malfunction
Summary report: N
TRANSPORT STRETCHER
MDR report key: 2243482
·
Received August 19, 2011
Report
- Report Number
- 3006697241-2011-00084
- Event Type
- Malfunction
- Date Received
- August 19, 2011
- Date of Event
- July 22, 2011
- Report Date
- July 22, 2011
- Manufacturer
- HILL-ROM DE MEXICO S DE RL DE CV
- Product Code
- FPO
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
REPAIRS HAVE NOT BEEN COMPLETED.
Description of Event or Problem · 1
INFO RECEIVED INDICATES THAT A NURSE AT (B)(6) HOSP SLIPPED AND FELL WHILE PUSHING A STRETCHER WITH A PT ON IT AND THE PT FELL OUT OF THE STRETCHER AS WELL. NO INJURY. THE ACCOUNT STATED THAT THE STEER CASTER ON THE STRETCHER WAS SHAKING, WHICH CAUSED THE NURSE TO LOSE CONTROL OF THE STRETCHER. THE STRETCHER WAS SHAKING SO MUCH THAT THE PT'S LEG STARTED TO FALL OFF THE STRETCHER. THE STEER CASTER WAS LOCKED OPPOSITE OF THE OTHER THREE CASTERS WHICH WERE IN THE TRAILING POSITION. THE POSITION OF THE STEER CASTER WAS THE CAUSE OF THE SHAKING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TRANSPORT STRETCHER | WHEELED STRETCHER | FPO | HILL-ROM DE MEXICO S DE RL DE CV | 8005 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |