HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1423500-2011-12071
- Event Type
- Malfunction
- Date Received
- September 12, 2011
- Date of Event
- August 24, 2011
- Report Date
- August 24, 2011
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE SYSTEM ERROR 2240 (AIR IN LINE) WAS NOT CONFIRMED DUE TO A LACK OF SAMPLE. THE ROOT CAUSE OF THE COMPLAINT WAS NOT DETERMINED. THE LOT NUMBER IS UNKNOWN; THEREFORE A BATCH REVIEW CANNOT BE PERFORMED. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.
(B)(4). A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF BAXTER'S INVESTIGATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.
A CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER REGARDING A SYSTEM ERROR (SE) 2240 (AIR IN LINE) ALARM THAT APPEARED ON THE HOMECHOICE (HC) DISPLAY DURING DWELL 4 OF 4. THE CALLER DID NOT SEE ANY OBVIOUS REASON FOR THE ALARM. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) EXPLAINED AND ASSISTED TO CLEAR THE ALARM. THE HP WOULD DISCARD THE SUPPLIES AND CONTACT THE NURSE REGARDING THE AIR DETECT ALARM AND BEING CONNECTED AS WELL AS ANY MISSED THERAPY. THERE WAS PATIENT INVOLVEMENT, BUT NO PATIENT INJURY OR MEDICAL INTERVENTION WAS INDICATED AT THE TIME OF THE INITIAL REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |