FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2243423 · Received September 12, 2011

Report

Report Number
2531779-2011-06720
Event Type
Malfunction
Date Received
September 12, 2011
Report Date
July 11, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(6) 2011 WITH THE FOLLOWING FINDINGS:EVALUATION REVEALED A DISLODGED DISPLAY SCREEN AND PARTIALLY DISLODGED FORCE SENSOR PINS. A REVIEW OF THE PUMP HISTORY INDICATED THAT LOSS OF CARTRIDGE DETECTION HAD OCCURRED WHICH COULD NOT BE DUPLICATED DURING TESTING. DURING EVALUATION THE PUMP WAS ABLE TO REWIND, LOAD AND PRIME SUCCESSFULLY.

Description of Event or Problem · 1

A FAMILY MEMBER REPORTED THAT THE PUMP WAS GIVING MULTIPLE LOSS OF PRIME WARNINGS. THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(6) 2011 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED A DISLODGED DISPLAY SCREEN AND PARTIALLY DISLODGED FORCE SENSOR PINS. THIS REPORT IS BEING MADE BASED ON EVALUATION RESULTS.