LAG SCREW, TI 10.5X95MM
Report
- Report Number
- 9610622-2011-00375
- Event Type
- Injury
- Date Received
- August 18, 2011
- Date of Event
- July 8, 2011
- Report Date
- July 26, 2011
- Manufacturer
- STRYKER OSTEOSYNTHESIS KIEL
- Product Code
- HSB
- PMA / PMN Number
- K032244
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- RISK MANAGER
Narratives
A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION.
PATIENT WAS ADMITTED TO THE HOSPITAL PRESENTING WITH EDEMA, SWOLLEN LEGS, AND FREQUENT FALLS. THE PATIENT HAS A HISTORY OF AN OPEN REDUCTION INTERNAL TROCHANTERIC NAIL FIXATION OF A LEFT HIP FRACTURE. THE NAIL WAS ORIGINALLY IN EXCELLENT POSITION. HOWEVER, THE HIP FRACTURE BEGAN TO COMPRESS OVER THE MONTHS AND THE LAG SCREW BEGAN TO MIGRATE LATERALLY AS THE HIP PROGRESSED. EXAMINATION FINDINGS WERE THAT THE LAG SCREW WAS SOMEWHAT PROMINENT. THE MD ADVISED THE PATIENT OF THE POSSIBLE NEED OF HARDWARE REVISION OR REPLACEMENT AND IT WAS POSSIBLE THAT OTHER RECONSTRUCTIVE PROCEDURES MIGHT NEED TO BE DONE IN THE FUTURE. THE MD ALSO GAVE THE PATIENT FOLLOW UP INSTRUCTIONS, BUT THE PATIENT HAD APPARENTLY MISSED SEVERAL VISITS AND HAD NOT BEEN SEEN IN THE OFFICE. HOWEVER, A RECENT CT OF HER PELVIS, PRESUMABLY TO EVALUATE THE SWELLING IN HER LEG, REVEALED THAT THE TROCHANTERIC NAIL REMAINS IN GOOD POSITION, THE DISTAL LOCKING SCREW REMAINS IN GOOD POSITION, BUT THE LAG SCREW HAS MIGRATED PROXIMALLY THROUGH THE TROCHANTERIC NAIL, AND IS NOW LODGED IN THE HEAD AND NECK OF THE FEMUR, TRAVERSING THE ACETABULUM, AND EXTENDS APPROXIMATELY 4 CM INTO THE ACETABULUM. THE PATIENT DENIES ANY BLOOD IN HER URINE OR ANY CHANGE IN HER BOWEL HABITS. THE DEVICE WAS IMPLANT IN (B)(6) 2010 AND EXPLANTED IN (B)(6) 2011. EXACT DATES WERE NOT PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LAG SCREW, TI 10.5X95MM | IMPLANT | HSB | STRYKER OSTEOSYNTHESIS KIEL | NA | K288284 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Other| R |