FDA Adverse Event Malfunction Summary report: N

COLLEAGUE

MDR report key: 2243355 · Received September 12, 2011

Report

Report Number
6000001-2011-23891
Event Type
Malfunction
Date Received
September 12, 2011
Date of Event
August 25, 2011
Report Date
September 1, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Removal / Correction Number
6000001-12/1/08-002-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ROOT CAUSE WAS DETERMINED TO BE USE/USER ERROR. THIS ISSUE IS BEING INVESTIGATED THROUGH CAPA. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED. A DEVICE HISTORY REVIEW WAS PERFORMED AND NO ABNORMALITY WAS OBSERVED IN THE MANUFACTURING OF THIS DEVICE. A SERVICE HISTORY REVIEW REVEALED THAT THE DEVICE WAS PREVIOUSLY SENT INTO SERVICE FOR THE REPORTED CONDITION. DURING PREVIOUS REPAIRS THE MAIN BATTERIES AND BATTERY HARNESS WERE REPLACED.

Additional Manufacturer Narrative · 1

(B)(4). THE REPORTED CONDITION OF DAMAGED BATTERY ALARM WAS CONFIRMED AND REPRODUCED DURING PRODUCT EVALUATION. THE ROOT CAUSE WAS IDENTIFIED AS DEPLETED MAIN BATTERIES. THE MAIN BATTERIES AND BATTERY HARNESS WERE REPLACED TO CORRECT THE REPORTED CONDTIION. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE REPORTED A COLLEAGUE INFUSION PUMP WITH A DAMAGED BATTERY ALARM. IT IS UNKNOWN WHEN THIS CONDITION OCCURRED. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. ACCORDING TO THE HOSPITAL REPRESENTATIVE, NO PATIENT INVOLVEMENT, INJURY, OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 6.13.90.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1