FDA Adverse Event Malfunction Summary report: N

INFUSOR

MDR report key: 2243346 · Received September 12, 2011

Report

Report Number
6000001-2011-23887
Event Type
Malfunction
Date Received
September 12, 2011
Date of Event
August 18, 2011
Report Date
August 18, 2011
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
MEB
PMA / PMN Number
K041738
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: ONE UNFILLED UNIT WAS RECEIVED BY BAXTER FOR EVALUATION. VISUAL EXAMINATION OF THE UNIT CONFIRMED THAT THE FILL PORT CAP WAS SEPARATED FROM THE FILL PORT. FURTHER EXAMINATION OF THE FILL PORT AND CAP SHOWED NO SIGNS OF PHYSICAL ABNORMALITY. THE ROOT CAUSE WAS OPERATOR ERROR DURING THE MANUAL FILL PORT CAP ASSEMBLY PROCESS. AS A RESULT OF THIS INCIDENT, AWARENESS TRAINING FOR ALL APPLICABLE MANUFACTURING OPERATORS WAS CONDUCTED IN AUGUST 2011. NO REPAIR WAS DONE, AS THIS IS A SINGLE-USE DEVICE WHICH WILL BE DISCARDED. NO OTHER OBSERVATIONS WERE NOTED ON THE UNIT. THE BATCH REVIEW REVEALED THAT ALL OF THE ACCEPTANCE CRITERIA WERE MET TO RELEASE THE LOT. THERE WERE NO NON-CONFORMANCES, FAILURES, REWORK, OR DEVIATIONS RELATED TO THE LOT. THERE WERE NO CHANGES TO SPECIFICATIONS, TEST METHODS, PROCESS, EQUIPMENT, OR RAW MATERIALS THAT COULD BE ASSOCIATED WITH THE REPORTED CONDITION. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Description of Event or Problem · 1

BAXTER AUSTRALIA RECEIVED A REPORT THAT AN INFUSOR HAD AN UNATTACHED LID. THIS CONDITION HAS THE POTENTIAL TO INTERRUPT THERAPY OR BREACH THE STERILE FLUID PATHWAY. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFUSOR PUMP, INFUSION, ELASTOMERIC MEB BAXTER HEALTHCARE - IRVINE 11E020

Patients

Seq Age Sex Outcome Treatment
1