Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO ZOLL MEDICAL CORPORATION; THE MALFUNCTION WAS DUPLICATED AND ATTRIBUTED TO A FAULTY LITHIUM BATTERY ON THE SYSTEM BOARD. ZOLL RECOMMENDS REPLACEMENT OF THE LITHIUM BATTERY EVERY 5 YEARS. ANALYSIS FOR REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WAS UNABLE TO POWER ON. COMPLAINANT INDICATED THAT THERE WAS NO PT INVOLVEMENT IN THE REPORTED MALFUNCTION.