FDA Adverse Event
Malfunction
Summary report: N
POSEY PERSONAL ALARM LOUD
MDR report key: 2243158
·
Received August 31, 2011
Report
- Report Number
- 2020362-2011-00288
- Event Type
- Malfunction
- Date Received
- August 31, 2011
- Report Date
- July 8, 2011
- Manufacturer
- J. T. POSEY CO.
- Product Code
- KMI
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). EVALUATION: RESULTS: EVALUATION OF THE RETURNED PRODUCT SHOWS THE ALARM DOES NOT SOUND WHEN IT SHOULD. THE EXTERNAL POWER SUPPLY SHOWS THE LOW BATTERY INDICATOR LIGHT DOES NOT COME ON WHEN IT SHOULD. THE 9V BATTERY SNAP HARD COVER IS BROKEN, BUT THE WIRES ARE STILL ATTACHED. NOTE: POSEY INSTRUCTION FOR USE WARNING STATEMENT: PERFORM SYSTEM TEST AND REMOVE THE MAGNET TO ACTIVATE THE ALARM. ALWAYS VERIFY YOU CAN HEAR THE ALARM VOLUME AT THE FURTHEST POSSIBLE DISTANCE BEFORE LEAVING THE PATIENT / RESIDENT UNATTENDED. (B)(4).
Description of Event or Problem · 1
CUSTOMER REPORTED THE ALARM HAS NO POWER. THE BATTERIES WERE REPLACED WITH A NEW SUPPLY. THERE IS NO VISIBLE DAMAGE TO THE OUTSIDE OF THE ALARM REPORTED. THERE WAS NO PT CONTACT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | POSEY PERSONAL ALARM LOUD | 8202L | KMI | J. T. POSEY CO. | 8202L | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |