FDA Adverse Event Injury Summary report: N

DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR

MDR report key: 2243035 · Received September 11, 2011

Report

Report Number
2017865-2011-06363
Event Type
Injury
Date Received
September 11, 2011
Date of Event
July 5, 2011
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WY
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. THE DAMAGE FOUND WAS SUSTAINED DURING THE SURGICAL PROCEDURE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT WAS EXPERIENCING SOME PAINFUL STIMULATION WITH PACING. THE LEAD WAS EXPLANTED WHEN REPOSITIONING WAS UNSUCCESSFUL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7120Q/65 NA

Patients

Seq Age Sex Outcome Treatment
1 65 YR Required Intervention