FDA Adverse Event Injury Summary report: N

DURATA STS OPTIM ACTIVE FIXATION

MDR report key: 2243001 · Received September 11, 2011

Report

Report Number
2017865-2011-06302
Event Type
Injury
Date Received
September 11, 2011
Date of Event
June 27, 2011
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. OTHER TEXT: DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN.

Additional Manufacturer Narrative · 1

LEAD FRACTURE WAS NOTED 27.1CM FROM THE CONNECTOR PIN. THE INNER COIL AND ONE OF THE SVC CABLES WERE FRACTURED.

Description of Event or Problem · 1

IT WAS REPORTED THAT NOISE WAS NOTED WITH ARM MOVEMENT. LEAD ISSUE SUSPECTED. SYSTEM WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DURATA STS OPTIM ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7120/65 NA

Patients

Seq Age Sex Outcome Treatment
1 18 YR Required Intervention