FDA Adverse Event Injury Summary report: N

RIATA ACTIVE FIXATION

MDR report key: 2242985 · Received September 11, 2011

Report

Report Number
2017865-2011-06078
Event Type
Injury
Date Received
September 11, 2011
Date of Event
July 21, 2011
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ANALYSIS FOUND INSULATION ABRASION.

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT OUT OF RANGE LOW IMPEDANCE WAS FOUND VIA (B)(4) TRANSMISSION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RIATA ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1580/65 NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention