FDA Adverse Event Malfunction Summary report: N

RIATA ACTIVE FIXATION

MDR report key: 2242982 · Received September 11, 2011

Report

Report Number
2017865-2011-06094
Event Type
Malfunction
Date Received
September 11, 2011
Date of Event
July 22, 2011
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Additional Manufacturer Narrative · 1

A PARTIAL LEAD WITH THE CONNECTOR PINS MEASURING 11.1CM WAS RETURNED FOR ANALYSIS. THE PORTION OF THE LEAD THAT WAS RETURNED WAS OTHERWISE NORMAL. WITHOUT RETURN OF THE ENTIRE LEAD A COMPLETE ANALYSIS COULD NOT BE PERFORMED.

Description of Event or Problem · 1

IT WAS REPORTED THAT AT FOLLOW-UP, DEVICE INTERROGATION FOUND EPISODES OF NOISE. PATIENT WILL BE MONITORED.

Description of Event or Problem · 1

NEW INFORMATION STATED THAT AT ROUTINE DEVICE CHANGE-OUT, HIGH IMPEDANCE WAS OBSERVED. LEAD WAS CAPPED AND A PORTION WAS RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RIATA ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1580/65 NA

Patients

Seq Age Sex Outcome Treatment
1 54 YR