CURRENT ACCEL VR, DF-4 CONNECTOR
MDR report key: 2242947
·
Received September 11, 2011
Report
- Report Number
- 2017865-2011-06532
- Event Type
- Injury
- Date Received
- September 11, 2011
- Date of Event
- June 2, 2011
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- LWS
- PMA / PMN Number
- P910023
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THE REPORTED PREMATURE BATTERY DEPLETION WAS VERIFIED IN THE LABORATORY. NO HIGH CURRENT DRAIN SOURCES WERE FOUND DURING BENCH, AUTOMATED ELECTRICAL, TEMPERATURE, AND HUMIDITY TESTING. THE BATTERY WAS SENT TO THE VENDOR FOR FURTHER ANALYSIS. AN INTERNAL BATTERY ANOMALY WAS FOUND TO CAUSE THE PREMATURE BATTERY DEPLETION.
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE CLINIC DUE TO A VIBRATORY ALERT. UPON INTERROGATION, PREMATURE BATTERY DEPLETION WAS OBSERVED. THE DEVICE WAS EXPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CURRENT ACCEL VR, DF-4 CONNECTOR | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | CD1215-36Q | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 18 YR | Required Intervention |