FDA Adverse Event Injury Summary report: N

CURRENT ACCEL VR, DF-4 CONNECTOR

MDR report key: 2242947 · Received September 11, 2011

Report

Report Number
2017865-2011-06532
Event Type
Injury
Date Received
September 11, 2011
Date of Event
June 2, 2011
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
LWS
PMA / PMN Number
P910023
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED PREMATURE BATTERY DEPLETION WAS VERIFIED IN THE LABORATORY. NO HIGH CURRENT DRAIN SOURCES WERE FOUND DURING BENCH, AUTOMATED ELECTRICAL, TEMPERATURE, AND HUMIDITY TESTING. THE BATTERY WAS SENT TO THE VENDOR FOR FURTHER ANALYSIS. AN INTERNAL BATTERY ANOMALY WAS FOUND TO CAUSE THE PREMATURE BATTERY DEPLETION.

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE CLINIC DUE TO A VIBRATORY ALERT. UPON INTERROGATION, PREMATURE BATTERY DEPLETION WAS OBSERVED. THE DEVICE WAS EXPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CURRENT ACCEL VR, DF-4 CONNECTOR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION CD1215-36Q NA

Patients

Seq Age Sex Outcome Treatment
1 18 YR Required Intervention