DURATA STS OPTIM ACTIVE FIXATION

MDR report key: 2242926 · Received September 11, 2011

Report

Report Number
2017865-2011-06440
Event Type
Malfunction
Date Received
September 11, 2011
Date of Event
July 29, 2011
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Additional Manufacturer Narrative · 1

A PARTIAL LEAD WITH THE PROXIMAL 26.5CM WAS RETURNED FOR ANALYSIS. EXTERNAL INSULATION ABRASION WAS NOTED AT 11.0-11.7CM FROM THE END OF THE CONNECTOR PIN, CONSISTENT WITH LEAD FRICTION TO THE ICD CAN. THE ETFE COATING WAS ABRADED AT THIS LOCATION. THIS IS CONSISTENT WITH THE OBSERVATION FROM THE FIELD OF NOISE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE HOLTER SHOWED EPISODES OF OVERSENSING, INCREASED CAPTURE THRESHOLDS AND DECREASED IMPEDANCE.

Description of Event or Problem · 1

NEW INFORMATION PROVIDED INDICATED THAT NOISE WAS OBSERVED ON ECG. THE LEAD WAS CAPPED. THE PATIENT IS IN GOOD CONDITION AFTER REPLACEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DURATA STS OPTIM ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7122/65 NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention