FDA Adverse Event Malfunction Summary report: N

DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR

MDR report key: 2242898 · Received September 11, 2011

Report

Report Number
2017865-2011-06386
Event Type
Malfunction
Date Received
September 11, 2011
Date of Event
July 28, 2011
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. (B)(4). NO COMPLAINT WAS RECEIVED WITH RETURN OF THE DEVICE. FAILURE (EVENT) OBSERVED DURING ANALYSIS. ANALYSIS FOUND INSULATION ABRASION ON THE OPTIM SHEATH AT 22.1CM TO 22.4CM FROM THE CONNECTOR PIN. HOWEVER, ONLY SURFACE ABRASION WAS NOTED ON THE SILICONE.

Description of Event or Problem · 1

THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.