FDA Adverse Event
Injury
Summary report: N
DURATA STS OPTIM ACTIVE FIXATION
MDR report key: 2242871
·
Received September 11, 2011
Report
- Report Number
- 2017865-2011-06303
- Event Type
- Injury
- Date Received
- September 11, 2011
- Date of Event
- June 9, 2011
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PATIENT RECEIVED MULTIPLE INAPPROPRIATE SHOCKS DUE TO NOISE ON THE RV LEAD. CHRONIC HIGH CAPTURE THRESHOLD WERE ALSO NOTED. FLUOROSCOPY REVEALED A FRACTURE CLOSE TO THE LEAD TIP. THE LEAD WAS CAPPED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DURATA STS OPTIM ACTIVE FIXATION | DEFIBRILLATION LEAD | NVY | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 7120/65 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR | Required Intervention |