FDA Adverse Event Injury Summary report: N

DURATA STS OPTIM ACTIVE FIXATION

MDR report key: 2242841 · Received September 11, 2011

Report

Report Number
2017865-2011-06285
Event Type
Injury
Date Received
September 11, 2011
Date of Event
July 7, 2011
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE THERAPY DUE TO OVERSENSING OF NOISE. NOISE WAS REPRODUCIBLE. A DECREASE IN SENSING WAS FOUND. THE LEAD WAS CAPPED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DURATA STS OPTIM ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7120/60 NA

Patients

Seq Age Sex Outcome Treatment
1 74 YR Required Intervention