FDA Adverse Event Injury Summary report: N

RIATA PASSIVE FIXATION

MDR report key: 2242775 · Received September 10, 2011

Report

Report Number
2017865-2011-06020
Event Type
Injury
Date Received
September 10, 2011
Date of Event
July 22, 2011
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE THERAPY DUE TO NOISE. NOISE WAS REPRODUCIBLE WITH ISOMETRICS. LOW IMPEDANCE WITH AN INCREASE IN CAPTURE WERE FOUND. LEAD FRACTURE SUSPECTED. AT DEVICE CHANGE OUT FOR ERI. LEAD WAS CUT, CAPPED, AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RIATA PASSIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1570/65 NA

Patients

Seq Age Sex Outcome Treatment
1 74 YR Required Intervention