FDA Adverse Event Malfunction Summary report: N

RIATA PASSIVE FIXATION

MDR report key: 2242772 · Received September 10, 2011

Report

Report Number
2017865-2011-06005
Event Type
Malfunction
Date Received
September 10, 2011
Date of Event
July 29, 2011
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT T-WAVE OVERSENSING WAS OBSERVED ON THE LEAD. MONITORING WILL CONTINUE. THE PHYSICIAN REPROGRAMMED THE DETECTION INTERVAL TO DELAY POSSIBLE INAPPROPRIATE THERAPY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RIATA PASSIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1560/60 NA

Patients

Seq Age Sex Outcome Treatment
1 77 YR