Report Number
2017865-2011-06225
Event Type
Injury
Date Received
September 10, 2011
Date of Event
June 29, 2011
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PACE/SENSE LEAD IMPEDANCE WAS MEASURING OUT OF RANGE. TRENDING SHOWED HIGH IMPEDANCES FOR THE LAST TWO YEARS. THE RV CAPTURE THRESHOLD WAS ALSO HIGH. LEAD FRACTURE SUSPECTED. THE PHYSICIAN CHOSE TO CAP THE LEAD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RIATA ST ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7000/65 NA

Patients

Seq Age Sex Outcome Treatment
1 71 YR Required Intervention