FDA Adverse Event Malfunction Summary report: N

RIATA ST ACTIVE FIXATION

MDR report key: 2242726 · Received September 10, 2011

Report

Report Number
2017865-2011-06249
Event Type
Malfunction
Date Received
September 10, 2011
Date of Event
June 1, 2011
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. (B)(4). NO COMPLAINT WAS RECEIVED WITH RETURN OF THE DEVICE. FAILURE (EVENT) OBSERVED DURING ANALYSIS. ANALYSIS OF THE LEAD FOUND INSULATION ABRASIONS AT 52.9CM-53.2CM AND 53.3CM-53.6CM FROM THE CONNECTOR PIN, CONSISTENT WITH FRICTION TO ANOTHER DEVICE.

Description of Event or Problem · 1

THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RIATA ST ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7011/65 NA