FDA Adverse Event Injury Summary report: N

RIATA ACTIVE FIXATION

MDR report key: 2242709 · Received September 10, 2011

Report

Report Number
2017865-2011-06173
Event Type
Injury
Date Received
September 10, 2011
Date of Event
June 9, 2011
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ONLY 53.7CM OF THE LEAD WAS RETURNED. INTERNAL ABRASIONS WERE FOUND AT TWO LOCATIONS BETWEEN 3.3CM TO 6.1CM FROM THE DISTAL TIP. THE RE CABLE AND THE INNER COIL WERE COMPROMISED.

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE THERAPY DUE TO NOISE. LEAD WAS EXPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RIATA ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1582/65 NA

Patients

Seq Age Sex Outcome Treatment
1 69 YR Required Intervention