RIATA ACTIVE FIXATION
Report
- Report Number
- 2017865-2011-06074
- Event Type
- Injury
- Date Received
- September 10, 2011
- Date of Event
- June 23, 2011
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
ONLY THE DISTAL PORTION OF LEAD WAS RETURNED, 54.9 CM LONG. ANALYSIS FOUND INTERNAL INSULATION ABRASION BETWEEN 13.6 CM AND 14.5 CM FROM THE DISTAL TIP. EXTERNAL INSULATION ABRASION WAS FOUND BETWEEN 44.6 CM AND 44.7 CM FROM THE DISTAL TIP; THIS DAMAGE IS CONSISTENT WITH FRICTION AGAINST THE ICD CAN. INTERNAL INSULATION ABRASIONS WERE FOUND BETWEEN 17.3 CM AND 19.0 CM FROM THE DISTAL TIP; ONE OF THE RV CONDUCTORS WAS COMPROMISED BETWEEN 18.5 CM AND 19.0 CM FROM THE DISTAL TIP.
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.
IT WAS REPORTED THAT THE PATIENT PRESENTED IN THE ER AFTER RECEIVING HV THERAPY. INTERROGATION FOUND AN ALERT FOR POSSIBLE HIGH VOLTAGE CIRCUIT DAMAGE WAS RECEIVED. LOW IMPEDANCE WAS OBSERVED AND CAPACITOR MAINTENANCE COULD NOT BE COMPLETED. HV THERAPY NOT AVAILABLE. LEAD FRACTURE SUSPECTED. LEAD WAS CAPPED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | RIATA ACTIVE FIXATION | DEFIBRILLATION LEAD | NVY | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 1580/65 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 86 YR | Required Intervention | (B)(4) |