FDA Adverse Event Injury Summary report: N

DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR

MDR report key: 2242519 · Received September 10, 2011

Report

Report Number
2017865-2011-06470
Event Type
Injury
Date Received
September 10, 2011
Date of Event
July 24, 2011
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE DAMAGE FOUND WAS SUSTAINED DURING THE SURGICAL PROCEDURE.

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT PRESENTED IN THE ER AFTER RECEIVING INAPPROPRIATE THERAPY DUE TO OVERSENSING. IT WAS FOUND THAT THE LEAD HAD DISLODGED. LEAD WAS EXPLANTED AND REPLACED WHEN REPOSITIONING WAS UNSUCCESSFUL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7122Q/58 NA

Patients

Seq Age Sex Outcome Treatment
1 29 YR Required Intervention