FDA Adverse Event
Injury
Summary report: N
DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR
MDR report key: 2242510
·
Received September 10, 2011
Report
- Report Number
- 2017865-2011-06330
- Event Type
- Injury
- Date Received
- September 10, 2011
- Date of Event
- June 29, 2011
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.
Description of Event or Problem · 1
THE PATIENT RECEIVED INAPPROPRIATE SHOCKS FOR SUSPECTED DOUBLE COUNTING. LOW R-WAVES AND VARYING CAPTURE THRESHOLDS WERE OBSERVED. LEAD DISLODGEMENT WAS NOTED. IT WAS REPORTED THAT THE PATIENT USED A WHEELBARROW TO MOVE BRICKS FIVE DAYS POST-IMPLANT. THE LEAD WAS REPOSITIONED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR | DEFIBRILLATION LEAD | NVY | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 7120Q/58 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Required Intervention |