FDA Adverse Event Injury Summary report: N

STYLE 15 SILICONE GEL FILLED BREAST IMPLANT

MDR report key: 22424869 · Received July 7, 2025

Report

Report Number
9617229-2025-11285
Event Type
Injury
Date Received
July 7, 2025
Date of Event
September 8, 2023
Report Date
October 9, 2025
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FTR
PMA / PMN Number
P020056
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN/WILL BE REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. THE REASON(S) FOR REOPERATION IS/ARE: RUPTURE.

Additional Manufacturer Narrative · 0

DEVICE ANALYSIS RESULTS: BASED ON THE PRODUCT ANALYSIS PERFORMED, THE ASSESSMENTS OF THE COMPLAINTS ARE: ¿ RUPTURE: OBSERVED BROKEN DEVICE ASSESSED AS FOLD FLAW OPENING. AS PER THE INVESTIGATION PROCEDURES, NON-PENETRATING NICKS WERE COMPLETED AND NONE OF THE OBSERVATIONS ARE FOUND TO BE POTENTIALLY RELATED TO THE MANUFACTURING PROCESS, NO FURTHER ACTIONS ARE REQUIRED.

Additional Manufacturer Narrative · 0

DHR TREND SUMMARY: THE REVIEW OF HISTORICAL COMPLAINTS ON THE COMPLAINT MANAGEMENT HANDLING INDICATES THAT THERE WERE NO OTHER RECORDS RELATED TO THIS EVENT FOR UNITS MANUFACTURED ON LOT NUMBER 2161746. THE SEARCH CRITERIA OF THESE QUERIES ARE MENTIONED ON QPP07-01- 004-HER1-G02 WORDING GUIDELINES FOR COMPLAINT INVESTIGATION CLOSURE REVISION 6.0 THE REVIEW OF ALL POTENTIAL TREND EVALUATIONS FOR BREAST IMPLANTS CLOSED DURING THE PAST 12 MONTHS INDICATES THAT NO CONFIRMED COMPLAINT TRENDS HAVE BEEN OBSERVED FOR THE EVENT A0412 MATERIAL RUPTURE. ADDITIONAL, CHANGED, AND/OR CORRECTED DATA: A2, B3, B5, B6, H6.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED "RUPTURE". THIS RECORD IS FOR THE LEFT SIDE. DEVICE REMAINS IMPLANTED.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED "RUPTURE" CONFIRMED VIA MRI. THIS RECORD IS FOR THE LEFT SIDE. THE DEVICE HAS BEEN EXPLANTED AND REPLACED.

Description of Event or Problem · 0

THE DEVICE HAS BEEN EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1194513 STYLE 15 SILICONE GEL FILLED BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR ALLERGAN (COSTA RICA) 2161746

Patients

Seq Age Sex Outcome Treatment
1 54 YR Female Required Intervention