FDA Adverse Event Injury Summary report: N

DURATA STS OPTIM ACTIVE FIXATION

MDR report key: 2242461 · Received September 10, 2011

Report

Report Number
2017865-2011-06458
Event Type
Injury
Date Received
September 10, 2011
Date of Event
June 21, 2011
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED ANALYSIS FOUND THAT THE SENSING COIL WAS PARTLY FRACTURED AT THE CRIMP SLEEVE TO THE CONNECTOR RING DUE TO FATIGUE. THIS COULD CAUSE THE REPORTED OVERSENSING.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE THERAPY DUE TO OVERSENSING ON THE LEAD. THE LEAD WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DURATA STS OPTIM ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7122/65 NA

Patients

Seq Age Sex Outcome Treatment
1 72 YR Required Intervention