HOMECHOICE
Report
- Report Number
- 1423500-2011-12036
- Event Type
- Malfunction
- Date Received
- September 10, 2011
- Date of Event
- June 1, 2011
- Report Date
- August 23, 2011
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED AND EVALUATED BY THE PRODUCT ANALYSIS LAB. THE PAL EVALUATED THE DEVICE AND NO FAILURE OR MALFUNCTION WAS IDENTIFIED THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED DIFFICULTY. THE DEVICE WAS TESTED AND PASSED THE HOMECHOICE RETURN INSTRUMENT TEST / EVALUATION (RITE) ELECTRICAL TEST BUT FAILED RITE FUNCTIONAL TEST. THE CAUSE OF THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) IDENTIFIED IN THE DEVICE LOG WAS DETERMINED TO BE UNDETERMINED. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.
(B)(4). THE DEVICE HAS BEEN RECEIVED, AND THE EVALUATION IS IN PROCESS. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.
DURING EVALUATION, ADDITIONAL ISSUES OF INCREASED INTRA PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED IN THE PATIENT THERAPY LOG ON (B)(6) 2011 AT 13:14 WITH ULTRAFILTRATION OF 232 ML DURING CYCLE 1. LARGEST PRESCRIBED FILL VOLUME: 2500ML.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |