FDA Adverse Event Malfunction Summary report: N

HOMECHOICE

MDR report key: 2242322 · Received September 10, 2011

Report

Report Number
1423500-2011-12034
Event Type
Malfunction
Date Received
September 10, 2011
Date of Event
August 22, 2011
Report Date
August 22, 2011
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED AND EVALUATED BY THE PRODUCT ANALYSIS LAB. AN EXTERNAL/INTERNAL INSPECTION WAS PERFORMED AND PASSED. THE ASSIGNABLE CAUSE FOR THE EARTH LEAKAGE CURRENT FAILED PERFORMANCE SPEC WAS DETERMINED TO BE A WET MACHINE.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED, AND THE EVALUATION IS IN PROCESS. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE (HC) DEVICE, A BAXTER TECHNICIAN DETERMINED THE HC MACHINE FAILED THE EARTH LEAKAGE CURRENT TEST.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1