FDA Adverse Event
Malfunction
Summary report: N
HOMECHOICE
MDR report key: 2242322
·
Received September 10, 2011
Report
- Report Number
- 1423500-2011-12034
- Event Type
- Malfunction
- Date Received
- September 10, 2011
- Date of Event
- August 22, 2011
- Report Date
- August 22, 2011
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED AND EVALUATED BY THE PRODUCT ANALYSIS LAB. AN EXTERNAL/INTERNAL INSPECTION WAS PERFORMED AND PASSED. THE ASSIGNABLE CAUSE FOR THE EARTH LEAKAGE CURRENT FAILED PERFORMANCE SPEC WAS DETERMINED TO BE A WET MACHINE.
Additional Manufacturer Narrative · 1
(B)(4). THE DEVICE HAS BEEN RECEIVED, AND THE EVALUATION IS IN PROCESS. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.
Description of Event or Problem · 1
DURING EVALUATION OF A RETURNED HOMECHOICE (HC) DEVICE, A BAXTER TECHNICIAN DETERMINED THE HC MACHINE FAILED THE EARTH LEAKAGE CURRENT TEST.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |