FDA Adverse Event
Malfunction
Summary report: N
LEGEND FOOTED ATTACHMENT
MDR report key: 2242319
·
Received September 10, 2011
Report
- Report Number
- 1625507-2011-00119
- Event Type
- Malfunction
- Date Received
- September 10, 2011
- Report Date
- August 12, 2011
- Manufacturer
- MDT POWERED SURGICAL SOLUTIONS
- Product Code
- HBB
- PMA / PMN Number
- K020069
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
REPORT CONFIRMED. EVALUATION DETERMINED THAT THE FOOTED PORTION WAS DAMAGED BY TOOL CONTACT. NO ADDITIONAL INFORMATION WAS AVAILABLE ON FOLLOW-UP. THE USER MANUAL CONTAINS THE FOLLOWING WARNING "THE ATTACHMENT SHOULD NOT BE USED IF ANY PART OF THE ATTACHMENT APPEARS TO BE BENT, LOOSE, MISSING, OR DAMAGED. EXCESSIVE PRESSURE OR IMPROPER HANDLING, SUCH AS BENDING OR PRYING, OF THE ATTACHMENT OR DISSECTING TOOL MAY CAUSE INJURY TO THE PATIENT, OPERATOR AND/OR OPERATING ROOM STAFF."
Description of Event or Problem · 1
DEVICE RETURNED FOR NON-SPECIFIED REPAIR. NO PATIENT IMPACT REPORTED. REPAIR ESCALATED TO COMPLAINT DUE TO ATTACHMENT DAMAGED BY TOOL. NO ADDITIONAL INFORMATION WAS AVAILABLE ON FOLLOW-UP.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LEGEND FOOTED ATTACHMENT | MOTOR, DRILL, PNEUMATIC | HBB | MDT POWERED SURGICAL SOLUTIONS | N/A | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |