FDA Adverse Event Malfunction Summary report: N

ACCESS® 2 IMMUNOASSAY SYSTEM

MDR report key: 2242296 · Received September 10, 2011

Report

Report Number
2122870-2011-03393
Event Type
Malfunction
Date Received
September 10, 2011
Date of Event
August 10, 2011
Report Date
August 10, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K922823/A007
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PER CUSTOMER, THE WASH BUFFER WAS SPILLED DUE TO A LEAK OF THE WASH BUFFER RESERVOIR FILTER SCREEN FITTING. SERVICE WAS NOT DISPATCHED. HOTLINE SENT THE CUSTOMER A NEW WASH BUFFER RESERVOIR WITH FITTING. HARDWARE IS THE ROOT CAUSE FOR THIS EVENT. BEC INTERNAL IDENTIFICATION (B)(4).

Description of Event or Problem · 1

THE CUSTOMER CONTACTED BECKMAN COULTER INC (BEC) TO REPORT THE FLUIDICS TRAY WAS LEAKING WASH BUFFER ONTO THE COUNTER BY THE ACCESS 2. 2. THE USER WAS WEARING PPE. NO REPORT OF INJURY OR USER EXPOSURE TO THE LIQUID WASTE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ACCESS® 2 IMMUNOASSAY SYSTEM DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER FOR CLINICAL USE JJE BECKMAN COULTER INC. ACCESS2 N/A