BECKMAN COULTER ¿ COULTER ® LH 750 HEMATOLOGY ANALYZER
Report
- Report Number
- 1061932-2011-01435
- Event Type
- Malfunction
- Date Received
- September 10, 2011
- Date of Event
- August 11, 2011
- Report Date
- August 11, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- GKZ
- PMA / PMN Number
- K011342
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
ON (B)(6) 2011, A BEC FIELD SERVICE ENGINEER (FSE) FOUND A LEAK AT THE LEE VALVE OF THE SLIDE MAKER ASPIRATION AREA. FSE REPLACED THE TUBING AND ALL REPAIRS WERE VERIFIED PER ESTABLISHED PROCEDURES. FSE INDICATED THAT THE LEAK WAS DUE TO A SPLIT IN THE DURO TUBING RUNNING THROUGH VL1 IN THE SLIDE MAKER ASPIRATION AREA. THE TUBING WAS SPLIT GOING THROUGH THE VALVE, DUE TO WEAR CAUSED BY NORMAL OPERATION. THE FSE ALSO STATED THAT ONLY DILUENT AND BLOOD LEAKED. ROOT CAUSE FOR THE LEAK IS THAT THE TUBING GOING THROUGH VL1 WAS SPLIT DUE TO WEAR. BEC INTERNAL IDENTIFICATION NUMBER IS (B)(4).
THE CUSTOMER CONTACTED BECKMAN COULTER INC (BEC) TO REPORT A LEAK AT THE SAMPLE ACCESS MODULE OF THE BECKMAN COULTER - COULTER LH 750 HEMATOLOGY ANALYZER. PER CUSTOMER, THE LEAK APPEARED TO BE BLOOD AND ISOTON. THE USER WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF GLOVES LAB COAT AND GLASSES AT THE TIME OF THE INCIDENT. THE MSDS WAS NOT REVIEWED, HOWEVER IT IS READILY AVAILABLE. MEDICAL ATTENTION WAS NOT SOUGHT. NO REPORT OF EXPOSURE TO MUCOUS MEMBRANES OR OPEN WOUNDS. THE LAB'S EXPOSURE CONTROL OR RISK MANAGEMENT PLANS ARE IN PLACE. NO DEATH, INJURY OR CHANGE TO PATIENT TREATMENT ATTRIBUTED OR ASSOCIATED TO THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BECKMAN COULTER ¿ COULTER ® LH 750 HEMATOLOGY ANALYZER | AUTOMATED DIFFERENTIAL CELL COUNTER | GKZ | BECKMAN COULTER INC. | LH750 | N/A |