UNICEL® DXC 800 SYNCHRON® SYSTEM

MDR report key: 2242272 · Received September 9, 2011

Report

Report Number
2050012-2011-05212
Event Type
Malfunction
Date Received
September 9, 2011
Date of Event
August 17, 2011
Report Date
August 23, 2011
Manufacturer
BECKMAN COULTER, INC.
Product Code
JJE
PMA / PMN Number
K042291
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

SERVICE WAS NOT DISPATCHED TO THE SITE TO ADDRESS THIS ISSUE. THE CUSTOMER SUSPECTED THE ISSUE WAS REAGENT RELATED. THE CUSTOMER HAS DEPLETED THEIR INVENTORY OF THE REAGENT LOT AND THE REAGENT LOT IS NO LONGER AVAILABLE IN THEIR LABORATORY. BECKMAN COULTER EVALUATION OF THE INSTRUMENT PERFORMANCE DATA DURING THE TIME OF THE EVENT INDICATED THAT HBA1C2 QUALITY CONTROL AND CALIBRATION RESULTS WERE WITHIN ESTABLISHED SPECIFICATIONS DURING THE TIMEFRAME OF THIS EVENT. THE CUSTOMER IS CURRENTLY IN THE PROCESS OF CONTACTING ADDITIONAL PATIENTS FOR REDRAWS AND RETESTS IN ASSOCIATION WITH THIS EVENT. A DEFINITIVE ROOT CAUSE FOR THIS EVENT IS UNKNOWN TO DATE. MDRS ASSOCIATED WITH THIS EVENT: 2050012-2011-05212, 2050012-2011-05213, 2050012-2011-05214, 2050012-2011-05215, 2050012-2011-05216.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT ERRONEOUS, ELEVATED HEMOGLOBIN A1C2 (HBA1C2) RESULTS WERE GENERATED FROM A UNICEL DXC 800 SYNCHRON SYSTEM, IN CONJUNCTION WITH A SPECIFIC HBA1C2 REAGENT LOT, FOR THIRTY PATIENTS OVER FIVE DAYS. THIS IS REPORT ONE OF FIVE AND REPRESENTS THE ERRONEOUS HBA1C2 RESULTS GENERATED FROM A UNICEL DXC 800 SYNCHRON SYSTEM FOR ONE PATIENT ON (B)(6) 2011. THE ERRONEOUS RESULT WAS REPORTED OUT OF THE LABORATORY AND WAS QUESTIONED BY A PHYSICIAN. THERE WAS NO DEATH, INJURY OR MODIFICATION TO PATIENT TREATMENT ASSOCIATED WITH THIS EVENT. UPON REPEAT TESTING ON THE SAME INSTRUMENT UTILIZING A DIFFERENT HBA1C2 REAGENT LOT, THE RESULT WAS LOWER AND CONSIDERED VALID. AN AMENDED REPORT WAS ISSUED.