COLLEAGUE
Report
- Report Number
- 6000001-2011-23849
- Event Type
- Malfunction
- Date Received
- September 9, 2011
- Date of Event
- August 16, 2011
- Report Date
- August 25, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
EVALUATION SUMMARY:THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH A DAMAGED BATTERY ALARM WAS CONFIRMED DURING PRODUCT EVALUATION. THE ROOT CAUSE OF THIS CONDITION WAS ASSIGNED TO DEPLETED MAIN BATTERIES RESULTING FROM USER ERROR. THE MAIN BATTERIES AND BATTERY HARNESS WERE REPLACED TO CORRECT THIS CONDITION.
(B)(4). THE DEVICE IS CURRENTLY IN THE PROCESS OF BEING EVALUATED ON SITE BY A BAXTER FIELD SERVICE TECHNICIAN. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.
THE FACILITY REPRESENTATIVE CONTACTED A TECHNICAL SERVICE REPRESENTATIVE TO REPORT A COLLEAGUE INFUSION PUMP WITH A DAMAGED BATTERY ALARM; THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. IT IS UNKNOWN WHEN THIS EVENT OCCURRED. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. THIS INVOLVED A REMEDIATED COLLEAGUE 2006 INFUSION PUMP WITH USER INTERFACE MODULE MASTER SOFTWARE VERSION 6.13.90. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |