FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER

MDR report key: 2242245 · Received September 9, 2011

Report

Report Number
2050012-2011-04958
Event Type
Malfunction
Date Received
September 9, 2011
Date of Event
August 10, 2011
Report Date
August 11, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HI, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER CLEANED THE REAGENT PROBE AND QC RECOVERED BETTER, BUT STILL RECOVERING ON THE LOW END. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON (B)(4) 2011 FOR THIS EVENT. THE FSE REPLACED THE CHEMISTRY CARTRIDGE (CC) SAMPLE PROBE AND PERFORMED REAGENT PROBE CARTRIDGE HEIGHT ALIGNMENT AND ROTARY ALIGNMENT. THE FSE TESTED OPERATION BY LOADING REAGENT PACKS WITH NO LEVEL SENSE ERRORS. CUSTOMER CALIBRATED TBIL AND RAN QC; ALL RESULTS WERE WITHIN ACCEPTABLE RANGE. THE ISSUE WAS RESOLVED.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) IN REGARDS TO OBTAINING SUPPRESSED ERRONEOUS TOTAL BILIRUBIN (TBIL) RESULTS GENERATED BY THE UNICEL DXC 600 PRO SYNCHRON CHEMISTRY ANALYZER. CUSTOMER ALSO INDICATED THAT QC HAS BEEN RECOVERING LOW. THE ERRRONEOUS RESULTS WERE REPORTED OUT OF THE LABORATORY. NO IMPACT TO PATIENT WAS REPORTED FOR THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER INC. DXC 600 PRO NA

Patients

Seq Age Sex Outcome Treatment
1