FDA Adverse Event Malfunction Summary report: N

IMMAGE® 800 IMMUNOCHEMISTRY SYSTEM

MDR report key: 2242239 · Received September 9, 2011

Report

Report Number
2050012-2011-05040
Event Type
Malfunction
Date Received
September 9, 2011
Date of Event
August 10, 2011
Report Date
August 10, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K962294
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

CUSTOMER TECHNICAL SUPPORT (CTS) ASSISTED THE CUSTOMER WITH COLD BOOTING THE SYSTEM. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON (B)(4) 2011 REPLACED A CLOGGED FILTER ON THE SAMPLE WASH CUP. REPAIRS WERE VERIFIED PER ESTABLISHED PROCEDURES PRIOR TO RETURNING IT INTO SERVICE. (B)(4).

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) STATING THAT THE SAMPLE WASH STATION WAS FLOODING IN THE IMMAGE 800 IMMUNOCHEMISTRY SYSTEM. NO OPERATOR EXPOSURE WAS REPORTED FOR THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 IMMAGE® 800 IMMUNOCHEMISTRY SYSTEM CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER INC. IMMAGE 800 N/A