IMMAGE® 800 IMMUNOCHEMISTRY SYSTEM
MDR report key: 2242239
·
Received September 9, 2011
Report
- Report Number
- 2050012-2011-05040
- Event Type
- Malfunction
- Date Received
- September 9, 2011
- Date of Event
- August 10, 2011
- Report Date
- August 10, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K962294
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
CUSTOMER TECHNICAL SUPPORT (CTS) ASSISTED THE CUSTOMER WITH COLD BOOTING THE SYSTEM. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON (B)(4) 2011 REPLACED A CLOGGED FILTER ON THE SAMPLE WASH CUP. REPAIRS WERE VERIFIED PER ESTABLISHED PROCEDURES PRIOR TO RETURNING IT INTO SERVICE. (B)(4).
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) STATING THAT THE SAMPLE WASH STATION WAS FLOODING IN THE IMMAGE 800 IMMUNOCHEMISTRY SYSTEM. NO OPERATOR EXPOSURE WAS REPORTED FOR THIS EVENT.