FDA Adverse Event Malfunction Summary report: N

INTERMATE

MDR report key: 2242231 · Received September 9, 2011

Report

Report Number
6000001-2011-23837
Event Type
Malfunction
Date Received
September 9, 2011
Date of Event
August 17, 2011
Report Date
August 17, 2011
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
FRN
PMA / PMN Number
K910425
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: BAXTER RECEIVED ONE SAMPLE FOR EVALUATION. THE REPORTED CONDITION OF A LEAK WAS CONFIRMED. UPON SAMPLE RECEIPT, THE DEVICE CONTAINED APPROXIMATELY 190 ML OF FLUID IN THE RESERVOIR. TO VERIFY THE REPORTED CONDITION, THE FILL-PORT CAP WAS REMOVED. UPON REMOVAL, LEAKAGE (BACKFLOW) WAS OBSERVED COMING OUT OF THE FILL PORT. VISUAL EXAMINATION OF THE DISASSEMBLED UNIT REVEALED THE ROOT CAUSE OF THE LEAK WAS A 1.1-MM PARTICLE OF ACRYLIC RESIN BENEATH THE CHECK BAND. NO REPAIR WAS DONE, AS THIS IS A SINGLE-USE DEVICE WHICH WILL BE DISCARDED. NO OTHER OBSERVATIONS WERE NOTED ON THE UNIT.

Additional Manufacturer Narrative · 1

(B)(4). THIS DEVICE IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES AND DOES NOT HAVE A 510(K) NUMBER. HOWEVER, THIS PRODUCT IS BEING REPORTED BECAUSE IT IS SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE U.S. A REQUEST FOR THE RETURN OF THE DEVICE HAS BEEN MADE. SHOULD THE DEVICE BE RECEIVED BY BAXTER FOR EVALUATION, A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF AN EVALUATION OR IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE. THE BATCH REVIEW REVEALED THAT ALL OF THE ACCEPTANCE CRITERIA WERE MET TO RELEASE THE LOT. THERE WERE NO NONCONFORMANCES, FAILURES, REWORK, OR DEVIATIONS RELATED TO THE LOT. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Description of Event or Problem · 1

BAXTER (B)(4) RECEIVED A REPORT THAT AN INTERMATE LEAKED AT THE FILL PORT DURING FILLING. THE DEVICE WAS BEING FILLED WITH A SOLUTION OF STERILE WATER. THIS CONDITION HAS THE POTENTIAL TO INTERRUPT THERAPY OR BREACH THE STERILE FLUID PATHWAY. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INTERMATE PUMP, INFUSION FRN BAXTER HEALTHCARE - IRVINE 09E074

Patients

Seq Age Sex Outcome Treatment
1