FDA Adverse Event Malfunction Summary report: N

INFUSOR

MDR report key: 2242215 · Received September 9, 2011

Report

Report Number
6000001-2011-23833
Event Type
Malfunction
Date Received
September 9, 2011
Date of Event
August 10, 2011
Report Date
August 17, 2011
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
MEB
PMA / PMN Number
K041738
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHARMACIST

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE IS AVAILABLE FOR EVALUATION PER THE CUSTOMER; HOWEVER, THE DEVICE HAS NOT YET BEEN RECEIVED BY BAXTER. SHOULD THE DEVICE BE RECEIVED BY BAXTER FOR EVALUATION, A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF AN EVALUATION OR IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: BAXTER RECEIVED ONE SAMPLE CONTAINING APPROXIMATELY 250 ML OF SOLUTION IN THE RESERVOIR. THE REPORTED CONDITION OF NO FLOW/NON-DELIVERY WAS NOT CONFIRMED. VISUAL EXAMINATION OF THE DEVICE SHOWED NO SIGNS OF BLOCKAGE THAT COULD HAVE CAUSED THE REPORTED CONDITION. FLOW WAS READILY OBSERVED AT THE LUER. FLOW CONTINUED WITHOUT STOPPING UNTIL THE SOLUTION WAS EMPTIED FROM THE RESERVOIR. NO ROOT CAUSE WAS IDENTIFIED, AS NO EVIDENCE OF PRODUCT MALFUNCTION WAS OBSERVED. NO REPAIR WAS DONE, AS THIS IS A SINGLE-USE DEVICE WHICH WILL BE DISCARDED. NO OTHER OBSERVATIONS WERE NOTED ON THE UNIT. THE BATCH REVIEW REVEALED THAT ALL OF THE ACCEPTANCE CRITERIA WERE MET TO RELEASE THE LOT. THERE WERE NO NON-CONFORMANCES, FAILURES, REWORK, OR DEVIATIONS RELATED TO THE LOT. THERE WERE NO CHANGES TO SPECIFICATIONS, TEST METHODS, PROCESS, EQUIPMENT, OR RAW MATERIALS THAT COULD BE ASSOCIATED WITH THE REPORTED CONDITION.

Description of Event or Problem · 1

BAXTER CANADA RECEIVED A REPORT THAT AN INFUSOR HAD NO FLOW AFTER FILLING. THE DEVICE WAS FILLED WITH A SOLUTION OF FLUOROURACIL. THIS CONDITION HAS THE POTENTIAL TO INTERRUPT THERAPY. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFUSOR PUMP, INFUSION, ELASTOMERIC MEB BAXTER HEALTHCARE - IRVINE 11A020

Patients

Seq Age Sex Outcome Treatment
1