FDA Adverse Event Malfunction Summary report: N

INTERMATE

MDR report key: 2242212 · Received September 9, 2011

Report

Report Number
6000001-2011-23827
Event Type
Malfunction
Date Received
September 9, 2011
Date of Event
July 19, 2011
Report Date
August 17, 2011
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
FRN
PMA / PMN Number
K910425
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: BAXTER RECEIVED ONE SAMPLE FOR EVALUATION. THE REPORTED CONDITION OF WHITE FLOW RESTRICTOR BREAKING OFF OF THE TUBING WAS NOT CONFIRMED. VISUAL INSPECTION OF THE UNIT DETERMINED THAT THE FLOW RESTRICTOR WAS COMPLETELY INTACT WITH THE TUBING WITHOUT ANY SIGNS OF DEFECT. NO REPAIR WAS DONE, AS THIS IS A SINGLE-USE DEVICE WHICH WILL BE DISCARDED. NO OTHER OBSERVATIONS WERE NOTED ON THE UNIT. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS.

Additional Manufacturer Narrative · 1

(B)(4). A REQUEST FOR THE RETURN OF THE DEVICE HAS BEEN MADE. SHOULD THE DEVICE BE RECEIVED BY BAXTER FOR EVALUATION, A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF AN EVALUATION OR IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE. THE BATCH REVIEW REVEALED THAT ALL OF THE ACCEPTANCE CRITERIA WERE MET TO RELEASE THE LOT. THERE WERE NO NONCONFORMANCES, FAILURES, REWORK, OR DEVIATIONS RELATED TO THE LOT.

Description of Event or Problem · 1

THE FACILITY REPORTED VIA PHONE THAT AN INTERMATE HAD THE WHITE FLOW RESTRICTOR BREAK COMPLETELY OFF OF THE TUBING BEFORE FILLING. THIS CONDITION HAS THE POTENTIAL TO INTERRUPT THERAPY OR BREACH THE STERILE FLUID PATHWAY. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INTERMATE PUMP, INFUSION FRN BAXTER HEALTHCARE - IRVINE 11E028

Patients

Seq Age Sex Outcome Treatment
1