FLEXICAP
Report
- Report Number
- 1423500-2011-12027
- Event Type
- Injury
- Date Received
- September 9, 2011
- Date of Event
- August 1, 2011
- Report Date
- August 24, 2011
- Manufacturer
- BAXTER HEALTHCARE - SWINFORD
- Product Code
- KDJ
- PMA / PMN Number
- K972579
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
(B)(4). AS THE DATE OF ONSET OF THIS PERITONITIS EPISODE IS UNKNOWN AND PATIENTS DISCARD SUPPLIES AFTER EACH THERAPY, THE SAMPLE WAS NOT REQUEST. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED. THIS IS REPORT 2 OF 4 INVOLVED IN THIS PERITONITIS EVENT.
(B)(4). A BATCH REVIEW WAS CONDUCTED ON POTENTIALLY ASSOCIATED LOT NUMBER 11B10H25 WITH NO DEFECTS NOTED. THE CAUSE OF THE PERITONITIS WAS USE ERROR-POOR ASEPTIC TECHNIQUE. THE LABEL REVIEW FOUND THE LABELING ADEQUATE FOR THE USE ERROR IN THIS COMPLAINT. BAXTER HAS RECEIVED SIMILAR REPORTS FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS.
THE FOLLOWING INFORMATION WAS RECEIVED FROM GLOBAL PHARMACOVIGILANCE (GPV) AND IS A SPONTANEOUS REPORT BY A CONSUMER WITH SUPPLEMENTAL INFORMATION BY A NURSE FROM THE USA OF PERITONITIS IN A PATIENT COINCIDENT WITH DIANEAL PD4 ULTRABAG THERAPY. DURING A CALL WITH BAXTER CUSTOMER SERVICES, THE FOLLOWING WAS REPORTED. ON AN UNREPORTED DATE IN 2011, THE PATIENT MADE MISTAKE/TOUCH CONTAMINATION/DID NOT WEAR A MASK. ON (B)(6) 2011, THE PATIENT EXPERIENCED PERITONITIS. THE PATIENT WAS HOSPITALIZED FROM (B)(6) 2011 FOR THE PERITONITIS. TREATMENT WAS NOT REPORTED. THE PATIENT WAS RECOVERING FROM THE PERITONITIS AND THE OUTCOME FOR MADE MISTAKE/TOUCH CONTAMINATION/DID NOT WEAR A MASK WAS NOT REPORTED. ON AN UNREPORTED DATE, DIANEAL THERAPY WAS WITHDRAWN. CONCOMITANT MEDICATIONS WERE NOT REPORTED. THE NURSE STATED THE PERITONITIS WAS NOT RELATED TO DIANEAL THERAPY AND WAS DUE TO THE PATIENT MADE MISTAKE/TOUCH CONTAMINATION/DID NOT WEAR A MASK. AN OPINION OF CAUSALITY WAS NOT REPORTED FOR THE PATIENT MADE MISTAKE/TOUCH CONTAMINATION/DID NOT WEAR A MASK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FLEXICAP | SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE | KDJ | BAXTER HEALTHCARE - SWINFORD |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 48 YR | Hospitalization | DIANEAL PD4 ULTRABAG |