COLLEAGUE
Report
- Report Number
- 6000001-2011-23757
- Event Type
- Malfunction
- Date Received
- September 9, 2011
- Date of Event
- August 22, 2011
- Report Date
- August 25, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS EVALUATED ON-SITE AT THE CUSTOMER FACILITY. BAXTER'S INVESTIGATION IS STILL IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.
(B)(4). THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH A DAMAGED BATTERY ALARM WAS CONFIRMED DURING EVALUATION. THE ASSIGNABLE CAUSE FOR THE REPORTED PROBLEM WAS DETERMINED TO BE DEPLETED BATTERIES RESULTING FROM USER ERROR. THE BATTERY AND BATTERY HARNESS WERE REPLACED TO FIX THE REPORTED CONDITION. A SERVICE HISTORY REVIEW REVEALED THAT THIS DEVICE WAS PREVIOUSLY SERVICED FOR THE REPORTED CONDITION. DURING PREVIOUS SERVICE, THE BATTERIES AND BATTERY HARNESS WAS REPLACED. A DEVICE HISTORY REVIEW WAS ALSO PERFORMED, FINDING THAT NO EXCEPTIONS WERE NOTED DURING THE MANUFACTURING OF THIS DEVICE.
THE FACILITY REPRESENTATIVE CONTACTED A TECHNICAL SERVICE REPRESENTATIVE TO REPORT A COLLEAGUE INFUSION PUMP WITH A DAMAGED BATTERY ALARM. THIS CONDITION HAS THE POTENTIAL TO CAUSE AN INTERRUPTION OF DELIVERY. IT IS UNKNOWN WHEN THIS EVENT OCCURRED. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. THIS INVOLVED A REMEDIATED COLLEAGUE 2006 INFUSION PUMP WITH USER INTERFACE MODULE MASTER SOFTWARE VERSION 6.13.90. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |