FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2242057 · Received September 9, 2011

Report

Report Number
2531779-2011-06707
Event Type
Malfunction
Date Received
September 9, 2011
Report Date
August 15, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: DURING EVALUATION CONTAMINATION WAS FOUND ON THE FORCE SENSOR AND THE FORCE SENSOR WAS FOUND TO BE OUT OF CALIBRATION. DURING TESTING, THE PUMP LOAD STEP WAS UNABLE TO DETECT THE FILLED CARTRIDGE.

Additional Manufacturer Narrative · 1

RECALL #: 2531779-03/24/2010/003-R. THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION IS NOT YET COMPLETE. WHEN EVALUATION IS COMPLETE, THE RESULTS WILL BE PROVIDED IN A SUPPLEMENTAL REPORT. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. THE OWNER'S BOOKLET STATES: "NEVER PRIME TUBING OR TIGHTEN THE CARTRIDGE CAP WHILE THE INFUSION SET IS CONNECTED TO YOUR BODY. DOING SO WHILE THE INFUSION SET IS CONNECTED TO YOUR BODY CAN RESULT IN UNINTENDED DELIVERY OF INSULIN, WHICH CAN RESULT IN SERIOUS INJURY OR DEATH."

Description of Event or Problem · 1

THE PATIENT REPORTED THAT HE WAS ATTACHED TO THE PUMP DURING A CARTRIDGE CHANGE AND THE PUMP DISPENSED INSULIN FROM THE CARTRIDGE DURING THE LOAD STEP. THE CARTRIDGE WAS FILLED WITH 200 UNITS OF INSULIN AND THE PATIENT REPORTS ABOUT ¾ OF THE INSULIN WAS EMPTIED INTO HIS BODY BEFORE HE PULLED THE SITE OUT OF HIS BODY. THE PATIENT REPORTEDLY LOST CONSCIOUSNESS AND HIS WIFE CALLED EMERGENCY SERVICES. THE PATIENT WAS TAKEN TO THE EMERGENCY ROOM AND HIS BLOOD GLUCOSE READING REGISTERED 20 MG/DL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANM 2020 INSULIN INFUSION PUMP

Patients

Seq Age Sex Outcome Treatment
1 55 YR Hospitalization| L| R