FDA Adverse Event Malfunction Summary report: N

NUCLEUS 24 CHANNEL COCHLEAR IMPLANT SYSTEM

MDR report key: 2242031 · Received September 9, 2011

Report

Report Number
6000034-2011-00685
Event Type
Malfunction
Date Received
September 9, 2011
Date of Event
August 17, 2011
Report Date
November 28, 2012
Manufacturer
COCHLEAR LTD.
Product Code
MCM
PMA / PMN Number
970051
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
AUDIOLOGIST

Narratives

Additional Manufacturer Narrative · 1

(B)(4): IMPLANTED DEVICE REMAINS.

Additional Manufacturer Narrative · 1

PER PATIENT'S SURGEON, THE DEVICE WAS EXPLANTED ON (B)(6) 2011; DURING THE SAME SURGERY THE PATIENT WAS REIMPLANTED WITH A NEW DEVICE. THIS REPORT IS FILED NOVEMBER 17, 2011.

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

PER THE CLINIC, THE PATIENT EXPERIENCED INTERMITTENCIES AND SUBSEQUENT LOSS OF CONNECTION TO THE INTERNAL DEVICE. REPROGRAMMING ATTEMPTS WERE MADE; HOWEVER, THE ISSUE COULD NOT BE RESOLVED. THE IMPLANTED DEVICE REMAINS. EXPLANT AND REIMPLANTATION IS PLANNED BUT HAS NOT TAKEN PLACE AT THE TIME OF THIS REPORT (B)(6), 2011.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NUCLEUS 24 CHANNEL COCHLEAR IMPLANT SYSTEM MCM, PRODUCT CODE: MCM MCM COCHLEAR LTD. CI512 N/A

Patients

Seq Age Sex Outcome Treatment
1 41 YR Required Intervention