FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2242016 · Received September 9, 2011

Report

Report Number
2531779-2011-06701
Event Type
Malfunction
Date Received
September 9, 2011
Report Date
August 10, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: REBOOTING WAS OBSERVED IN THE BLACK BOX HISTORY. NO DAMAGE WAS OBSERVED TO THE BATTERY COMPARTMENT OR CAP. THE BATTERY CAP SECURED APPROPRIATELY TO THE PUMP AND THE PUMP POWERED UP NORMALLY. THE PUMP WAS EXERCISED FOR 24 HOURS WITHOUT POWER ISSUES OR ALARMS OCCURING. THE BATTERY CAP WAS TESTED AND NO CONTACT DEFECTS WERE FOUND. THE PUMP WAS OPENED AND MOISTURE DAMAGE WAS FOUND ON THE PCB.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

THE PATIENT CLAIMED THAT THERE IS AN INTERMITTENT POWER ISSUE. REPORTEDLY, THE ANIMAS PUMP STARTED TO REBOOT SEVERAL TIME ON (B)(6) 2011. THE PATIENT STATED THAT THE ANIMAS PUMP FELL DURING PATIENT'S CHEERLEADING EVENT. THERE WAS NO PRODUCT MISUSE. THE BATTERY CAP AND COMPARTMENT WAS NOT DAMAGED. THE BATTERY COMPARTMENT WAS CORROSION FREE. THE BATTERY WAS REPLACED BUT THE ISSUE WAS NOT RESOLVED WITH TRAINING. THE PRODUCT WAS SENT BACK FOR INVESTIGATION. THE PATIENT WAS ADVISED TO GO ON THE BACKUP PLAN. THERE WAS NO REPORT OF ANY PATIENT IMPACT ASSOCIATED WITH THE COMPLAINT. THIS COMPLAINT IS BEING REPORTED AS A MALFUNCTION DUE TO THE POWER ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 16 YR