HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1423500-2011-12016
- Event Type
- Malfunction
- Date Received
- September 9, 2011
- Date of Event
- August 22, 2011
- Report Date
- August 22, 2011
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THERE WAS NO ALLEGATION REPORTED AGAINST THE BAXTER PRODUCT BY THE CUSTOMER; THEREFORE, THE SAMPLE WAS NOT REQUESTED FOR EVALUATION AND A BATCH REVIEW WILL NOT BE CONDUCTED. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.
(B)(4). BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH CAPA: RENQ-CAPA-(B)(4). THE LABEL REVIEW FOUND THE LABELING ADEQUATE FOR THE USE ERROR IN THIS COMPLAINT. THE COMPLAINT WAS CONFIRMED, AND THE ROOT CAUSE WAS DETERMINED TO BE THE PATIENT NOT BEING CONNECTED WHEN THERAPY INITIATED.
A CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER REGARDING THE HOME CHOICE (HC) SYSTEM ERROR 2240 (AIR IN LINE), THIS SYSTEM ERROR OCCURRED DURING INITIAL DRAIN. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) ASSISTED THE HOME PATIENT (HP) BY EXPLAINING THE MESSAGE TO THE HP AND HAD HER RECYCLE THE MACHINE. THE PATIENT WAS INVOLVED BUT THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 41 YR | HOME CHOICE |