FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2241902 · Received September 9, 2011

Report

Report Number
1423500-2011-11997
Event Type
Malfunction
Date Received
September 9, 2011
Date of Event
March 25, 2011
Report Date
August 23, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). SYSTEM ERROR 2240 (AIR IN LINE) WAS CONFIRMED AS IT WAS IDENTIFIED IN THE ALARM LOGS OF THE RETURNED DEVICE. THE ROOT CAUSE WAS UNDETERMINED. A BATCH REVIEW WAS NOT PERFORMED AS THE LOT NUMBER WAS UNKNOWN. BAXTER HAS FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. ADDITIONAL INVESTIGATION IS BEING CONDUCTED THROUGH CAPA (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). THERE WAS NO ALLEGATION OF A PRODUCT MALFUNCTION. THEREFORE, A BATCH REVIEW AND SAMPLE EVALUATION WILL NOT BE CONDUCTED. A FOLLOW-UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION IS OBTAINED.

Description of Event or Problem · 1

A SYSTEM ERROR (SE) 2240 (AIR IN LINE) ALARM WAS IDENTIFIED IN THE LOG OF A RETURNED HOMECHOICE DEVICE. THE SYSTEM ERROR OCCURRED ON (B)(6) 2011. IT IS UNKNOWN IF THIS ALARM OCCURRED DURING PATIENT THERAPY, HOWEVER, NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.