FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2241852 · Received September 9, 2011

Report

Report Number
2531779-2011-06692
Event Type
Malfunction
Date Received
September 9, 2011
Report Date
August 14, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 09/01/2011 WITH THE FOLLOWING FINDINGS: ALL KEYPAD BUTTONS WERE FOUND TO BE RESPONDING INTERMITTENTLY. THE KEYPAD WAS REMOVED AND ADHESIVE WAS OBSERVED UNDER THE BUTTON CONTACTS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

THE PATIENT REPORTED THAT THE BUTTONS ARE ALL RESPONDING INTERMITTENTLY WHEN PRESSED. THE PATIENT INDICATED THAT THE 'OK' BUTTON IS UNRESPONSIVE AND DOES NOT SPRING BACK WHEN PRESSED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 21 YR