DA VINCI SURGICAL SYSTEM

MDR report key: 2241809 · Received September 9, 2011

Report

Report Number
2955842-2011-00293
Event Type
Other
Date Received
September 9, 2011
Date of Event
August 11, 2011
Report Date
August 11, 2011
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K070684
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION CONDUCTED BY THE FIELD SERVICE ENGINEER CONCLUDED THAT THE SYSTEM ERROR CODES EXPERIENCED BY THE SITE WAS ASSOCIATED WITH A PATIENT SIDE MANIPULATOR (PSM). THE PSM IS AN INSTRUMENT ARM LOCATED ON THE PATIENT SIDE CART AND PROVIDES THE STERILE INTERFACE FOR THE ENDOWRIST INSTRUMENTS. THE SYSTEM WAS REPAIRED BY REPLACING THE AFFECTED PSM. SYSTEM ERROR CODES 20008 AND 21008 APPEARS WHEN THE DAVINCI SAFETY SYSTEM, THE ANGULAR POSITION OF ONE OR MORE ROBOTIC JOINTS ON THE SPECIFIED MANIPULATOR, AS MEASURED BY THAT JOINT'S PRIMARY CONTROL SENSOR (ENCODER) AND THE SECONDARY SENSOR (POTENTIOMETER), WERE OUT OF SPECIFIED TOLERANCE FOR AGREEMENT. THE PSM WAS RETURNED FOR FAILURE ANALYSIS INVESTIGATION. ENGINEERING EVALUATION OBSERVED THAT THE AXIS 7 CABLES HAD BECOME DERAILED FROM ITS PULLEYS. THE PSM WAS REPAIRED BY REPLACING THE CABLES. AS OF (B)(4) 2011, THERE HAVE BEEN NO REPORTED RECURRENCES OF THE ISSUE AT THIS HOSPITAL.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI PROSTATECTOMY PROCEDURE, THE SURGICAL STAFF EXPERIENCED SYSTEM ERROR CODE 20008 AND 21008. THEY SITE ATTEMPTED TO TROUBLESHOOT THE ISSUE BY RESTARTING THE SYSTEM AND APPLYING EMERGENCY POWER OFF (EPO), HOWEVER, THE SYSTEM ERROR CODE PERSISTED. AT THIS TIME, THE SURGEON MADE THE DECISION TO CONVERT THE PLANNED SURGICAL PROCEDURE TO A TRADITIONAL OPEN SURGICAL TECHNIQUES. NO PATIENT HARM WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DA VINCI SURGICAL SYSTEM ENDOSCOPIC INSTRUMENT CONTROL SYSTEM NAY INTUITIVE SURGICAL,INC. IS1200 A4.3P9

Patients

Seq Age Sex Outcome Treatment
1 65 YR Other DA VINCI SYSTEM INSTRUMENTS AND ACCESSORIES